1、Main technical parameters high frequency
(1)Power requirements
Single-phase power supply: 220V ± 22V safety standard socket
Power frequency :50Hz±1Hz
battery capacity:4kVA
Power supply resistance:<0.5Ω
(2)Standard sizes:
The highest distance from the ground: 1800mm±20mm
The ball is at least the distance from the ground: 490mm±20mm
Equipment parking size: 1400×700×1330(mm)
Equipment quality: 130(kg)
(3)The main technical parameters
Rated output power: 3.2 kW
Tube XD6-1.1、3.5/100(Fixed anode tube XD6-1.1, 3.5 / 100)
Anode target angle 19°
Limiter manual adjustment
Fixed filter 2.5mm 2.5mm aluminum equivalent X-ray tube with beam restraint
Positioning lights halogen bulb; average illuminance of not less than 100 Lx
/1 m SID: 430 mm × 430 mm Maximum cartridge size / 1 m SID
The maximum floor slope when moving ≤10°
Rated output power 3.5kW100kV×35mA=3.5kW
Tube voltage(kV):40~110kV
Tube current(mA):30~70 mA
Exposure time (s):0.04~5 s
Current and tube voltage regulation range
4、Features:
For hospital wards and emergency room photography dedicated
Flexible mobile operating performance
Wireless remote exposure, greatly reducing the dose of physician radiation
5、Detail image:
E-mail: service@newheek.com
Company: Weifang Newheek Electronic Technology Co., Ltd.
ADD: E Building of Future Star Scientific Innovation Industrial Zone of No.957 Wolong East Street, Yulong Community, Xincheng Sub-District Office, Weifang Hi-tech Zone, Shandong Province, China