Power requirements:
Single-phase power supply: 220V±22V (safety standard socket)
Power frequency: 50Hz±1Hz
Power capacity: 4kVA
Power supply internal resistance: <0.5Ω
Specifications and dimensions:
Tube horizontal extension: ≥800mm
The maximum distance between the tube and the ground: 1800mm±20mm
The minimum distance between the tube and the ground: 490mm±20mm
Equipment parking size: 1400×640×1300(mm)
Equipment mass: 130(kg)
Main technical parameters:
Rated output power: 3.2 kW
Tube: Fixed anode tube XD6-1.1, 3.5/100
Anode target angle: 19°
Beam Limiter: Manual Adjustment
Fixed filter: 2.5mm aluminum equivalent (X-ray tube with beam limiter)
Positioning light: 24V halogen bulb; average illuminance not less than 100 Lx
Maximum cassette size/1 m SID: 450 mm × 450 mm
Maximum ground slope when moving ≤10°
Rated output power: 3.2kW (100kV×32mA=3.2kW
Tube voltage (kV): 40~100kV (1kV increment/decrement)
Tube current (mA): 32 mA, 40 mA, 50 mA
Exposure time (s): 0.01~4 s
Current time product (mAs): 1~200mAs
●For photography in hospital wards and emergency rooms
●Flexible mobile operation performance
●Wireless remote control exposure, greatly reducing the radiation dose of doctors